Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
42

8.CALIBRATIONMODE
This chapter describes the functions available when the CALIBRATION mode is selected on the
transmitter.
Forspecificinformationaboutcalibratingthetransmitterseechapter10,InstrumentCalibration.
The CALIBRATION mode is accessed bypressing andholding theSETUP keyfor approximately 4
secondsuntilthewords“CalibrationMenu”appearsatthebottomofthedisplay.Thetransmitterwill
returntotheRUNmodewhentheSETUPkeyispressedagainor60secondsafterthelastkeypress.
ChangingoptionsintheCALIBRATIONmodeisthesameastheSETUPmode.Seechapter6.2Changing
Options.
8.1FUNCTIONSUMMARYTABLE
WhenthetransmitterisintheCALIBRATIONmodetheSETUPlightwillbeonandthewords“Commissioning
Menu”willbeatthebottomofthedisplay.Thefollowingtableshowsthe
CALIBRATIONmenufunctions:
Menu
#
Functionname Range Default
value
1 ReferenceVoltage1,50mV 40.00to60.00mV 47.14mV
2 ReferenceVoltage2,200mV 180.00to210.00mV 182.24mV
3 ReferenceVoltage3,1200mV 1150.0to1250.0mV 1221.8mV
4 ReferenceVoltage4,2500mV 2400.0to2550.0mV 2489.2mV
5 OutputChannel1,Calibration AutoCalibratedManualCalibratedSet
4mASet20mA
Auto
Calibrated
6 OutputChannel1,4mATrim 3.00to5.00mA 4.00mA
7 OutputChannel1,20mATrim 19.00to21.00mA 20.00mA
8 OutputChannel2,Calibration AutoCalibratedManualCalibratedSet
4mASet20mA
Auto
Calibrated
9 OutputChannel2,4mATrim 3.00to5.00mA 4.00mA
10 OutputChannel2,20mATrim 19.00to21.00mA 20.00mA
11 AmbientTemperatureSensorOffset 10.0°Cto10.0°C 0.0°C
12 LowOxygenCal1 80.0%to120.0% 100.00%
13 LowOxygenCal2 80.0%to120.0% 100.00%
14 TransmitterOutputSelect 420mA/020mA/420mAover
scalelimit/020mAoverscalelimit
420mA
15 TransmitterOutputLimiting Disabled/Hold0mA/Hold4mA/
Hold20mA
Hold0mA
16 MainsVoltageDetectionOverride Automatic/220/240/10/120 Automatic
17 MainsFrequencyDetectionOverride Automatic/50Hz/60Hz Automatic
18 HeaterSSRSelectionNormal Heater1<>Heater2Heaters<>Cal
Purge
Normal
19 SSRFailProtection Enabled/Disabled Enabled