Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
36

7.2COMMISSIONINGMODEFUNCTIONS
7.2.1InternalDate/Time
Thedate andtime areusedin thetransmitter toruntime basedoperationssuchas impedance
readings and gas calibrationchecks.Thecorrect date and timeshould be entered bypressingand
holdingtheOPTIONUP&DOWNkeys.ThedateandtimewillchangefasterthelongertheOPTION
keyisheld.
7.2.2ServiceDate
Theservicedatecanbe usedtokeepa recordofwhena probewaschanged.TheRUNTIMEtimer
whichkeepsatrackofthehoursandminutesthatthetransmitterandprobehasbeenoperatingis
resetwhentheservicedateischanged.
TheservicedateandtheruntimetimercanbedisplayedaslowerlineitemsinRUNmode.
7.2.3NumberofProbes
The9060Ztransmittercanoperateinsingleordualprobemode.Byselectingsingleprobe,alldual
probeoptionsandalarmswillbedisabledandsecondprobeoptionshiddenfromtheuser.
Ifyouareusingthetransmitter asa singleprobe deviceyou shouldset thenumber ofprobesto
‘singleprobe’todisableallprobe2alarmsanddisplayitems.
7.2.4Probe1&2Type
Theprobetype functionallowsaselectionbetweenaheated probeandan unheatedprobetypes.
Heatercontrolandalarmswillbedisabledifunheatedprobetypeisselected.
7.2.5Probe1&2ThermocoupleType
The thermocouple type can be set to K, J, R or S. The thermocouple in the Teledyne Analytical
Instruments9060HheatedprobesarealwaysKtype.9060UL/UHunheatedprobesareavailableinR
typeorStype.Otherthermocoupleoptionsaremadeavailableforspecialinstallations.
7.2.6AuxiliaryThermocoupleType
Whenthetransmitterisoperatingasasingleprobemode,theprobe2thermocoupleinput(terminals
7&8)isusedasanauxiliarythermocoupleinput.Indualprobemodethisoptionisunavailable.
Useofanauxiliarythermocoupleallowsforbasicmonitoringoftemperatureaswellasenablingthe
calculationofcombustionefficiency.ThethermocoupletypecanbesettoK,J,RorS.
7.2.7TransmitterOutputChannel1&2
The9060Ztransmitterhastwofullyconfigurable420mAanalogueoutputs.Thechannelscanbe
configuredindependentlytooutputoneofseveralcalculatedvalues.
