Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
24

networklinefromthetransmittersistakenfromthemiddleofthetransmitternetworkstring,a
terminatingresistorshouldbeenabledwithLK3ateachendofthenetworkline.
The MODBUS™ protocol of the network is –
BaudRate 19,200
Parity Even
StopBits 1
RS485 2wire(halfduplex)
Mode RTU(binarymode)
FormoredetailsseeAppendix5.


4.14ONNECTINGPOWER
Beforecommissioningtheprobe,sensorortransmitter,readtheCAUTIONparagraphsatthefrontof
thismanual.Thegreen&yellowEARTHwireMUSTbeconnectedtoearth.
4.15COMMISSIONING‐RUNMODE
When the transmitter is turned on it will go to RUN mode. The SETUP/RUN button will toggle
between the two modes. Theupperline ofthe display will nowread ‘% OXYGEN’. Ifthe probe or
sensortemperatureisnotabove650°C(1200°F),a“ProbeLowTemperature”messageisflashedon
thelowerline.Theprobeorsensortemperaturecanbecheckedonthelowerlineofthedisplay.
4.16PROBEORSENSORCALIBRATION
The zirconia sensor provides an absolute measurementof oxygen partial pressure. There are no
calibrationadjustments,apartfrom‘Probe1Offset’,fortheprobe.ThezirconiasensorEMFiseither
correctoritneedstobereplaced.Tocheckthattheprobeisfunctioningcorrectly,firstcheckthatthe