Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
43

8.2CALIBRATIONMODEFUNCTIONS
8.2.1REFERENCEVOLTAGES
Thecalibrationoftheanalogueinputsisbasedonthevoltageofatemperaturecompensatedvoltage
referenceintegratedcircuit.There are4voltagesgeneratedfromthestandardreference voltage.
Theywillvarybyabout1%fromonetransmittertoanotherbutcanbetrimmedbysettingtheactual
voltagesintoCALIBRATION menufunctions #1to4.Thesevoltagesshould bemeasured andthe
referencevoltageentriesinthemenucheckedeveryyear.
8.2.2OUTPUTCHANNEL1AND2CALIBRATION
Theoutput420mAcalibration canbe seteither AUTOMATICALLY(default)or MANUALLY.If Auto
Calibratedisselectedthetransmitterwillgothroughanoutputcalibrationcyclewhenthepoweris
turnedonorwhentheAUTOCALbuttonispressedinSETUPmode.Thiswilldiverttheoutputsback
totheinputandautomaticallysetthe4mAand20mAcalibration.
IfManually Calibratedis selectedthenthe 4mA and20mAcalibration must beset inthe nexttwo
functions.Forthefullexplanationseechapter10.1.2.
8.2.3AMBIENTTEMPERATURECALIBRATION
The ambient temperature measurement is used as the cold junction temperature for the
thermocouplemeasurements.Useanindependenttemperaturesensortomeasurethetemperature
inside the transmitter case near the screwterminal #1. Enter this temperature into the calibration
function11byusingtheOPTIONkeysandthentheENTERkey.
8.2.4LOWOXYGENCALIBRATION
Thelowoxygen calibrationfactors canbe usedto finetune theoxygen calculationat lowoxygen
readings.Itwillnotaffectthemeasurementat20.9%.Theyareincludedtoallowoxygenprobesmade
byothermanufacturerstobeusedonaTeledyneAnalyticalInstrumentstransmitter.Thedefaultis
100%butcanbesettobetween80and120%.
8.2.5TRANSMITTEROUTPUTSCALE
Thetransmitteroutputscanbeconfiguredtoeither420mAor020mA.Thetwoscaledoutputsare
capable of outputting 024mA and will transmit overscale unless the overscale limit option is
selected.
8.2.6TRANSMITTEROUTPUTLIMITING
Iftheoxygenreadingisinvalidtheoutputcanbemadetobesetto0mA,4mAor20mA.Aninvalid
readingiswhentheprobetemperatureisbelow650°C(1200°F)ortheprobethermocouplehasbeen
detectedasopen circuit.Thedefault istoset theoutput to20mAbut thefunction canalsobe
disabled.
8.2.7MAINSVOLTAGEDETECTION
Thedefaultsettingforthe9060Ztransmitteristhatitwillautomaticallydetectthemainsvoltageand
frequencyinordertocorrectlydrivetheprobeheaters.Ifthereisanyuncertaintyinthisdetection
systemthevoltagecanbeforcedto220240or100127,andthefrequencycanbesetto50or60Hz.