Probeinsertionlength(fromprocessendofmountingthreadtoprobesensingtip.
3.1.3SERIES9060HPROBESSPECIFICATIONS
MODEL 9060H 9060UL/UH
Application Combustionflue Combustionfluegases
gasesbelow above700°C(1290°F)with
900°C(1650°F) nocontaminants.
Refertonote1 eg.naturalgas,lightoils
TemperatureRange 0to900°C.Refertonote2
(32to1650°F)
700to1400°C
(1470to2550°F)
Length 250to2000mm 500to1500mm
(10”to80”) (20”to60”)
ProcessConnection 1½BSP"orNPT ¾BSP" orNPT
ElectricalConnection Weatherproof plug‐in connector or optional screw terminals.The plug
connectorcanbesuppliedwiththecable.
Cable Orderaspecificlengthwiththetransmitter
Heater Yes No
Thermocouple K,integral R,integral
ResponseTime Typically<4secs. Typically<1sec
HeadTemperature ‐25to100°C(‐15to210°F)withweatherproofconnector
‐25to150°C(‐15to300°F)withscrewterminals
ReferenceGas Ambient air, 50 to 150 cc/min (3 to 9 scim). Pump can be supplied with
transmitter.
RefAirConnection 1/4NPT" Integral air line in probe cable.
Barbedfittingto3/16"IDPVCtube.
Filter Removable sintered titanium alloy particulate filter, 30 micron standard,
optional15micronavailable.Refertonote2
CalibrationCheckGas
Connection
1/8NPTfemale" 1/8NPTfemale"
Weight 2kg(4.4lb)plus
165g(5.8oz)/100mm
(4”)length
1kg(2.2lb)plus
100g(3.5oz)/100mm
(4”)length
Notes: 1. Care must be taken to avoid contact with explosiveor inflammablegases with 9060H
heated probes when hot. Teledyne Analytical Instruments transmitters have built in safety
protection.
2.Processgastemperaturemustbebelow800°Cifthefiltersarefitted.
Pleasecontactfactoryforcorrosivesotherthansulfur orzinc.Wecanprovidetestmaterialstotry
inyouratmosphere.