Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
38

7.2.8FluePressureUnitsandValue
The9060Ztransmitteriscapableoffixedorvariableinputpressurecompensationinthecalculationof
oxygenfrom‐1Atmto+3Atm.Iftheprobeisrunninginapressurizedenvironmentthisvalueshould
besettoallowforaccurateoxygenmeasurement.Ifthefluepressureisconstantselect'Fixed"inthis
functionandsetthepressureunitsandvalueintheproceedingmenus.Ifthepressurevaries,select
"Variable",andconnectapressuretransducertoterminals24&25.(Seesection4.23).Settherange
ofthetransducerusingazeroandspanvalueinthecommissioningmenus.
7.2.9TemperatureUnits
ThedisplayonthetransmittercanbechangedtoshowtemperatureineitherCelsiusorFahrenheit
scales.
7.2.10CalibrationFreezesOutputs
Duringgascalibrationchecksitmayberequiredthattheanalogueoutputsremainfrozenattheirlast
reading.Whencalfreezesisenabledbothanalogueoutputsremainfrozenuntiltheentirecalibration
cyclehascompleted.
7.2.11Solenoid1&2Operation
The twosolenoidoutputs can be individually configured to perform gas calibration check or probe
purge.
Someof themenusbetween functions22 to28(31 to37) willnot appeardependingon whathas
beenselectedinfunction#20and#21(29and30).
7.2.12Solenoid1&2Automatic/Manual
Whensettoautomaticthesolenoidwillperformapreprogrammedgas calibrationorpurgecycle.
Whensettomanualthesolenoidiscontrolleddirectlyfromthecorrespondingkeyonthefrontofthe
case.Whensettoautomaticthecalibrationcheckorthepurgecyclecanbestartedbypressingthe
correspondingkeyonthefrontofthecase.Itcanbestoppedbypressingthesamekeyagain.
7.2.13Solenoid1&2StartTime
Forautomaticgascalibrationcheckingorpurgeeventsastarttimeisspecifiedtoallowprecisecontrol
overthetimingofsuchevents.Startingatthisspecifiedtime,thenextgascalibrationorpurgeevent
willoccurinsuccessiveintervalsfromthistime.
7.2.14Solenoid1&2Period
Thisoptionspecifiestheperiodbetweenautomaticgascalibrationorpurgeevents.Forperiodsless
than24hourstheperiodisdivisibleinto24hoursforcingschedulingofeventstooccuratthesame
timeeachdaybetween1minuteand7days.
The ‘NoTimed Operation’ option accessible by scrollingthe option below 1min
ute allowsforpre
configured solenoid cycles to be triggered manually via the keypad, but without the automatic
scheduling.
7.2.15Solenoid1&2Duration
Durationoftimethatanautomaticgascalibrationorpurgeeventenergizesthesolenoid