Technical Alarm Messages (INOPs) 4 Patient Alarms and INOPs
75
tcpO
2
(or tcpCO
2
or tcGas)
EQUIP MALF
Numeric is replaced by -?-
INOP tone.
tcGas There is a malfunction in the transducer or module. Connect
another transducer. If this INOP persists, contact your service
personnel.
tcpO
2
(or tcpCO
2
or tcGas)
NO TRANSDUC
Numeric is replaced by -?-
INOP tone.
tcGas No transducer is connected to the tcpO
2
/tcpCO
2
module.
Silencing the alarm switches off the measurement.
tcpO
2
(or tcpCO
2
or tcGas)
STABILIZING
Numeric is displayed with a ?
tcGas The transducer has not yet reached the selected temperature
and/or skin hyperemization is not yet finished. This INOP will
disappear within three minutes.
tcpO
2
(or tcpCO
2
or tcGas)
UNPLUGGED
Numeric is replaced by -?-
INOP tone.
tcGas The measurement is switched on but the module is unplugged.
The measurement automatically disappears from the display.
Silencing this INOP switches off the measurement.
<∆Temp> CHK SOURCES
Numeric is replaced by -?-
TEMP
Difference
Not all measurements or values required to perform the
calculation are available. Check measurement sources.
<∆Temp> CHK UNITS
Numeric is replaced by -?-
TEMP
Difference
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
<Temp label> DEACTIVATED
INOP tone
TEMP A Temp measurement label in the measurement server or
extension has been deactivated, either by connecting a Pressure
transducer in the shared Press/Temp socket, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp
transducer or reactivate the measurement label in the
Measurement Selection window.
<Temp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
TEMP Contact your service personnel.
The temperature hardware is faulty.
<Temp label> NO
TRANSDUCER
Numeric is replaced by -?-
INOP tone.
TEMP Make sure the TEMP probe is connected to the MMS or
module.
If you silence this INOP, the measurement will be switched off.
<Temp label> UNPLUGGED
INOP tone
TEMP A Temp measurement label has been deactivated, either by
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
<Temp label> OVERRANGE
Numeric is replaced by -?-
INOP tone.
TEMP Try changing the application site of the transducer.
[The temperature is less than -1
°C, or greater than 45°C.]
Tesoph INOPS TEMP See <Temp label> INOPs (under Temp).
TimeExpired:<timer label>
INOP tone
Monitor The time has expired for the timer indicated in the INOP text.
Clearing the timer clears the INOP.
Tnaso INOPS TEMP See <Temp label> INOPs (under Temp).
Trect INOPS TEMP See <Temp label> INOPs (under Temp).
Tskin INOPS TEMP See <Temp label> INOPs (under Temp).
INOP Message, Indication Source What to do