Philips MP60/70/80/90 Medical Alarms User Manual


 
4 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
62
BISx INCOMPATIBLE
INOP tone
BIS The BISx software is not compatible with the BIS module or
with the MP20/MP30 monitor software. A software upgrade
may be required. Contact your service personnel.
BISx MALFUNCTION
INOP tone
BIS The BISx is faulty. Disconnect and reconnect it to the module
or BIS interface board. If the INOP persists, replace the BISx.
MP20/MP30 - Malfunction on interface board. If the INOP
persists, contact your service personnel.
CANNOT ANALYZE ECG ECG/
Arrhythmia
The arrhythmia algorithm cannot reliably analyze the ECG
data. Check the ECG signal quality of the selected primary and
secondary leads. If necessary, improve lead position or reduce
patient motion.
If you have arrhythmia analysis on, and you are not getting a
reliable HR because the signal is below a minimum amplitude,
unstable, or contains artifact, and you have tried to improve the
system performance by choosing another lead and changing
electrodes, you should consider turning arrhythmia analysis off.
CANNOT ANALYZE ST ST The ST algorithm cannot generate a valid ST value. Possible
causes are large variations in the measured ST values for
consecutive beats, or ventricular paced beats. Review the ECG
signal quality and the ST measurement points.
If the patient has a ventricular pacemaker, ST analysis is not
possible.
CCI NO BSA
CCI numeric unavailable
INOP tone
C.O. CCI cannot be calculated because the patient's body surface
area is unknown. Enter the patient weight and height to
provide the BSA for CCI calculation.
CCO BAD PRESS SIGN
numeric is replaced by -?-
INOP tone
C.O. The arterial pressure wave can currently not be used for pulse
contour calculation for CCO or CCI measurement. Possible
causes are air bubbles in the tubing or a physiological
condition, for example severe arrhythmia.
CCO NO <Pressure label>
numeric is replaced by -?-
INOP tone may sound
C.O. CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO menu under CCO From
matches the pressure measured with the arterial catheter for
CCO measurement. A VueLink pressure cannot be used. Select
ano the r pr es sur e lab el, ei the r AB P, Ao , ART, B AP, FAP, or UAP.
CCO NO CALIBRATION
numeric is replaced by -?-
C.O. The CCO measurement is currently not calibrated.
CCO NO PRESS
at Information Center
C.O. CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO menu under CCO From
matches the pressure measured with the arterial catheter for
CCO measurement. A VueLink pressure cannot be used. Select
ano the r pr es sur e lab el, ei the r AB P, Ao , ART, B AP, FAP, or UAP.
CCO NOT SUPPORTED
numeric is replaced by -?-
INOP tone
C.O. A catheter for transpulmonary C.O. measurements has been
unplugged and replaced with a Right Heart C.O. catheter, or
the measurement mode has been changed manually. Silencing
this INOP switches the measurement off.
CCO/CCI OVERRANGE
numeric is replaced by -?-
INOP tone
C.O. The measured CCO or CCI value is not within the specified
range for CCO/CCI measurement.
CCO <Pressure label>
INVALID
numeric is replaced by -?-
INOP tone may sound
C.O. The arterial pressure selected for pulse contour calculation for
CCO is available but currently invalid. Make sure the pressure
transducer is connected and the zero calibration is valid.
INOP Message, Indication Source What to do