Philips MP60/70/80/90 Medical Alarms User Manual


 
Measurement Specifications 34 Installation and Specifications
351
Measurement Validation: The SpO
2
accuracy has been validated in human studies against arterial
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be expected to fall within the specified
accuracy compared to CO-oximeter measurements. Display Update Period: Typical: 2 seconds,
Maximum: 30 seconds. Max. with NBP INOP suppression on: 60 seconds.
SpO
2
Performance Specifications
SpO
2
The specified accuracy
is the root-mean-
square (RMS)
difference between the
measured values and
the reference values
Range
0 to 100%
Option #A01
Accuracy
Philips Reusable Sensors:
M1191A, M1191AL, M1192A, M1192AN = 2% (70% to 100%)
M1193A, M1194A, M1195A, = 3% (70% to 100%)
M1191T, M1192T, M1193T (Adult) = 3% (70% to 100%)
M1193T (Neonate) = 4% (70% to 100%)
Philips Disposable Sensors with M1943A(L):
M1132A, M1133A (adult/infant) = 2%
M1901B, M1902B, M1903B, M1904B, M1131A, M1133A (neonate)
= 3% (70% to 100%)
NellcorPB
®
Sensors with M1943A(L):
MAX-A, MAX-AL, MAX-P, MAX-I, MAX-N, D-25, D-20, I-20, N-
25, OxiCliq A, P, I, N = 3% (70% to 100%)
Masimo Sensors
®
with AC-1:
LNOP Adt, Adtx,Pdt, Pdtx, INF-L, DC-I, DC-IP, Y-I: 2% (70% to
100%)
LNOP NEO-L, NEOPT-L: 3% (70% to 100%)
LNOP TC-I: 3.5% (70% to 100%)
Option #A02
Accuracy
Philips Reusable Sensors
M1191ANL, M1192AN = 2% (70% to 100%)
M1193AN, M1194AN, M1195AN = 3% (70% to 100%)
Philips Disposable Sensors with M1943NL:
M1901B (Adult), M1902B, M1903B, M1904B = 2% (70% to 100%)
M1901B (Neonate) = 3%
NellcorPB
®
Sensors with M1943NL:
MAX-FAST, MAX-A, MAX-AL, MAX-P, MAX-I, MAX-N (Adult) =
2% (70 - 100%)
SC-A, OxiCliq A, P, I, N (Adult) = 2.5% (70 - 100%)
MAX-N (Neonate), Dura-Y D-YS (Infant to Adult), DS-
100A,Oxiband OXI-A/N (Adult), OXI-P/I = 3% (70 - 100%)
MAX-R, OxiCliq N (Neonate), SC-NEO (Neonate), SC-PR
(Neonate), Dura-Y D-YS with D-YSE or D-YSPD clip = 3.5% (70 -
100%)
Dura-Y D-YS (Neonate), Oxiband OXI-A/N (Neonate) = 4% (70 -
100%)
Resolution
1%