Philips MP60/70/80/90 Medical Alarms User Manual


 
Monitor Safety Specifications 34 Installation and Specifications
339
Environmental Specifications
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective range
for the combined products is that range which is common to the specifications for all products.
Monitor MP40, MP50, MP60, MP70, MP80, MP90
Item Condition Range
Temperature Range Operating 0 to 35 °C (32 to 95 °F)
Non-operating
-20 to 60 °C (-4 to 140 °F)
Battery storage
-20 to 50 °C (-4 to 122 °F)
Monitor MP60, MP70 in conjunction with IntelliVue 802.11 Bedside Adapter (option #J35)
Temperature Range Operating 0 to 30
o
C (32 to 86
o
F)
Monitor MP20, MP30
Item Condition Range
Temperature Range Operating 0 to 40 °C (32 to 104°F)
Operating when
charging battery or
when equipped with
IntelliVue Instrument
Telemetry
0 to 35 °C (32 to 95 °F)
Non-operating -20 to 60 °C (-4 to 140 °F)
Battery storage
-20 to 50 °C (-4 to 122 °F)
Monitor MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90
Item Condition Range
Humidity Range Operating 20% to 85% Relative Humidity (RH) (non condensing)
Non-operating
5% to 85% Relative Humidity (RH)
Altitude Range Operating -500 m to 3000 m (10000 ft)
Non-operating
-500 m to 12000 m (40000 ft)