Philips MP60/70/80/90 Medical Alarms User Manual


 
23 Event Surveillance Viewing Events
250
Event bars: Each event bar represents one event. The height indicates the event severity. Bars that
extend over more than one channel represent combi-events. Manually-captured events are marked with
a bar above the timeline instead of in the measurement channels.
Event values: Event information for the currently-selected event is shown on the left of the review
window. The trigger measurement is highlighted.
If an alarm triggers the event, the monitor shows the alarm conditions that triggered the event.
If the event trigger was user-defined, instead of an alarm condition the monitor shows for
example
SpO2 94<96, where the second number is the current event trigger threshold and the
first number is the maximum deviation from the set limit.
Trigger threshold: The horizontal lines show the trigger thresholds. Gaps in the line indicate that the
trigger was inactive for a while, possibly because alarms were switched off or because there was an
INOP condition in the group measurements.
Timeline: The timeline at the bottom indicates the period currently stored in the event database.
Event Episode Window
To enter the Event Episode window, select the pop-up key Show Episode.
Depending on the event group settings, the
Event Episode window shows either 20 minutes of
average trend event information, four minutes of high-resolution event information, or 15 seconds of
realtime wave information.
HR
Event Review
(Total: 4)
SpO2
Resp
Manual
Total
Brady
DESAT
APNEA