Philips MP60/70/80/90 Medical Alarms User Manual


 
Resolving Patient Information Mismatch 5 Managing Patients
83
Patient Mismatch - If One Set of Patient Data is Correct
If there is a mismatch between an
Information Center and a monitor,
choose the data set you want to
continue using for this patient,
either:
Continue Central: to continue with
the patient demographics from the
Information Center, discharge the
patient in the monitor, and use the
default monitor profile.
Continue Monitor: to continue with
the patient in the monitor and
discharge the patient in the
Information Center, permanently
deleting all data in the Information
Center.
If there is a mismatch between a
monitor and a measurement server,
choose the data set you want to
continue using for this patient,
either:
Continue Monitor: to continue with
the patient demographics, trend
data, and settings in the monitor.
This discharges the patient in the
measurement server, and resets all
MMS settings to the defaults
currently active for the monitor.
Continue MMS: to upload the data
- patient demographics, trend data
(if configured), and measurement
settings (if configured) - stored in the MMS to the monitor. This clears all data in the monitor,
resets the monitor to the default Profile, and discharges the patient in the monitor.
Patient Mismatch - If Neither Patient Data Set is Correct
A patient mismatch where neither set of patient data is correct might occur if you connect a new MMS
to a monitor in order to prepare for a new patient, before you actually start measuring.
Select New Patient if you are sure that none of the information is correct. This discharges all
patients, erases all data in both the monitor and MMS, resets all settings to the default Profile, and
lets you admit a new patient.
Patient Selection
Central Monitor
Last name DOE MILLER
First name JOHN
MRN 1234HG9556
Patient Cat Adult Neo
Paced No Yes
Continue
Central
Continue
Monitor
New
Patient
Same
Patient
Patient Selection
Monitor MeasServ
Last name MILLER ADAMS
First name PETER
MRN
Patient Cat Neo Neo
Paced Yes No
Continue
Monitor
Continue
MMS
New
Patient
Same
Patient