Fluke 1595A Thermometer User Manual


 
1594A/1595A Super-Thermometer
Interface Commands
78
Query form: CALC:AVER:COUN?
Alternate parameter: DEF, MIN, MAX
Sets the statistics Sample Count setting.
Parameter <value> is the maximum number of samples in the moving window when the Statistics Mode set-
ting is LIMITED (see Section 8.7.3.22, CALC:AVER:TCON NORM|MOV, on page 79). The range is 1
through 100 and the default is 100.
The *RST command sets Sample Count to 100.
8.7.3.16 CALC<n>:AVER:MAX?
Example command: CALC3:AVER:MAX?
Example response: 231.928833
Reads a channel’s statistical maximum (see Section 5.7.4.2, FIELD SETTINGS (F2), on page 48).
Sufx <n> is the channel number: 1 through 4 for the corresponding front channel input, 5 through 14 for
channels of the rst attached scanner, and 15 through 24 for channels of the second attached scanner.
The returned value has the same units as the measurement returned by the FETC? command. If the calculation
is temperature, the unit of measurement is determined by the Unit setting (see Section 5.7.4.5, TEMPERA-
TURE UNIT (F5), on page 50).
If no result is available, no response is returned and error -230,“Data corrupt or stale” is generated in the error
queue.
8.7.3.17 CALC<n>:AVER:MIN?
Example command: CALC3:AVER:MIN?
Example response: 231.928207
Reads a channel’s statistical minimum (see Section 5.7.4.2, FIELD SETTINGS (F2), on page 48).
Sufx <n> is the channel number: 1 through 4 for the corresponding front channel input, 5 through 14 for
channels of the rst attached scanner, and 15 through 24 for channels of the second attached scanner.
The returned value has the same units as the measurement returned by the FETC? command. If the calculation
is temperature, the unit of measurement is determined by the Unit setting (see Section 5.7.4.5, TEMPERA-
TURE UNIT (F5), on page 50).
If no result is available, no response is returned and error -230,“Data corrupt or stale” is generated in the error
queue.
8.7.3.18 CALC<n>:AVER:SDEV?
Example command: CALC3:AVER:SDEV?
Example response: 0.000144
Reads a channel’s statistical sample standard deviation (see Section 5.7.4.2, FIELD SETTINGS (F2), on page
48).
Sufx <n> is the channel number: 1 through 4 for the corresponding front channel input, 5 through 14 for
channels of the rst attached scanner, and 15 through 24 for channels of the second attached scanner.
The returned value has the same units as the measurement returned by the FETC? command. If the calculation
is temperature, the unit of measurement is determined by the Unit setting (see Section 5.7.4.5, TEMPERA-
TURE UNIT (F5), on page 50).
The standard deviation is available if at least four measurements have been acquired from the channel. If
no result is available, no response is returned and error -230,“Data corrupt or stale” is generated in the error
queue.