Fluke 1595A Thermometer User Manual


 
95
Remote Operation
Interface Commands
8.7.3.95 SYST:BEEP[:IMM]
Example command: SYST:BEEP
Produces an immediate beep from the speaker.
8.7.3.96 SYST:BEEP:KEY <boolean>
Query form: SYST:KEY:BEEP?
Alternate parameter: DEF
Sets the Key Beep setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default setting is
ON.
8.7.3.97 SYST:BEEP:ALER <boolean>
Query form: SYST:KEY:ALER?
Alternate parameter: DEF
Sets the Alert Beep setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default setting is
ON.
8.7.3.98 SYST:CODE:VERS?
Example response: 1.23,1.00,1.10
Reads the rmware code version numbers. The response contains the following:
<UI version>,<kernel version>,<analog version>
8.7.3.99 SYST:COMM:GPIB:ADDR <value>
Query form: SYST:COMM:GPIB:ADDR?
Alternate parameter: DEF, MIN, MAX
Sets the IEEE-488 port’s Address setting (see Section 8.6, Conguring the IEEE-488 Port, on page 69).
8.7.3.100 SYST:COMM:GPIB[:STAT] <boolean>
Query form: SYST:COMM:GPIB?
Sets the IEEE-488 port’s Port Enable setting (see Section 8.6, Conguring the IEEE-488 Port, on page 69).
8.7.3.101 SYST:COMM:GPIB:TERM LIN|RET
Query form: SYST:COMM:GPIB:TERM?
Sets the IEEE-488 port’s Termination setting (see Section 8.6, Conguring the IEEE-488 Port, on page 69).
8.7.3.102 SYST:COMM:SER:BAUD <value>
Query form: SYST:COMM:SER:BAUD?
Sets the serial port Baud Rate setting (see Section 8.5, Conguring the RS-232 Serial Port, on page 69).
8.7.3.103 SYST:COMM:SER:LIN <boolean>
Query form: SYST:COMM:SER:LIN?
Alternate parameter: DEF
Sets the serial port’s Linefeed setting (see Section 8.5, Conguring the RS-232 Serial Port, on page 69).
8.7.3.104 SYST:COMM:SOCK:ADDR <IP address>
Example command: SYST:COMM:SOCK:ADDR “147.120.0.11”
Query form: SYST:COMM:SOCK:ADDR?