Fluke 1595A Thermometer User Manual


 
1594A/1595A Super-Thermometer
Main Menu
32
5.7.1.2.1.1.3 EDIT RESISTOR (F2)
MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | DEFINE RESISTOR |
The Edit Resistor function key opens the highlighted Rs denition for editing. The conguration elds are
explained in the Add Resistor section above.
5.7.1.2.1.2 READ RESISTOR (F3)
MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | DEFINE RESISTOR |
Read Resistor allows the user to load a pre-congured reference resistor denition from a USB memory
device.
To load a resistor denition, insert a USB memory device that contains a resistor conguration into the front
panel. Allow about 10 seconds for the system to recognize the USB device and then press OK (F5). A list of
resistor denition les stored on the memory device appears on the screen. Scroll through the list using the
Up/Down arrow keys and select the le to be loaded. Press OK (F5) to begin the le transfer. A message will
prompt the user to continue.
Tip: Resistor denition les must be located in a folder named “\159x\resistor” on the USB memory
device. The les must have the extension “.res” and contain resistor denition data in the proper format.
This type of le can be created by using the Write File function, see Section 5.7.1.2.1.3.2, WRITE FILE
(F2), on page 32 in the Manage Resistors menu.
5.7.1.2.1.3 MANAGE RESISTORS (F3)
MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs |
Rs resistor denitions in the resistor denition library may be added, deleted, or prioritized in the Manage
Resistors menu using the following function keys.
5.7.1.2.1.3.1 READ FILE (F1)
MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | MANAGE RESISTORS
The Read File function key allows the user to import an entire set of resistor denitions from a le on a USB
memory device.
Every resistor denition in the resistor denition library must have a unique ID. If a resistor denition in the
library has the same ID as one read from the USB memory device, it will be replaced.
Tip: Resistor denition les must be located in a folder named “\159x\resistor” on the USB memory
device. The les must have the extension “.res” and contain resistor denition data in the proper format.
This type of le can be created by using the Write File function, see Section 5.7.1.2.1.3.2, WRITE FILE
(F2), on page 32 in the Manage Resistors menu.
5.7.1.2.1.3.2 WRITE FILE (F2)
MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | MANAGE RESISTORS
The Write File function key in the Manage Resistors menu allows the user to export one resistor denition or
the entire library to a le on a USB memory device.
If a resistor was selected in the Manage Resistors screen, that resistor denition will be written to a le. If ALL
RESISTORS was selected, the entire Resistor Library will be written to a le.
The screen shows the name of the le to be written. The user may change the name of the le by pressing
ENTER to open the alpha-numeric editing screen. Finally, the user presses the CONTINUE function key to