5.7.2.5.3 MOVE UP (F3)
MAIN MENU | PROBE MENU | MANAGE PROBES |
The Move Up function key changes the order of the probe denitions in the Probe Library by moving the
selected denition up one position until it reaches the top of the list.
5.7.2.5.4 MOVE DOWN (F4)
MAIN MENU | PROBE MENU | MANAGE PROBES |
The Move Down function key changes the order of the probe denitions in the Probe Library by moving the
selected denition down one position until it reaches the bottom of the list (before ALL PROBES).
5.7.2.5.5 DELETE PROBE (F5)
MAIN MENU | PROBE MENU | MANAGE PROBES |
The Delete Probe function key allows the user to delete probe denitions from the Probe Library. If the Probe
Library capacity of 100 denitions has been reached, a probe denition has to be deleted before adding an-
other denition.
A single highlighted probe denition can be highlighted and then deleted or all probe denitions can be delet-
ed if ALL PROBES is selected. A message prompts the user to conrm the delete action before it takes place.
5.7.3 MEASURE MENU (F3)
MAIN MENU |
Figure 12 Measure Menu Screen
The Measure Menu contains the function keys needed for setting up channel scanning, measurement timing,
digital lter settings, controlling measurement recording, and performing zero-power measurements.