Remote Operation
Interface Commands
Password protection: Measure Protection
Sets the Average Enable for the sample average (see Section 5.7.3.2, TIMING SETTINGS (F2), on page
39).
The *RST command sets Average Enable to the default setting which is ON.
8.7.3.84 SENS:AVER2[:STAT] <boolean>
Query form: SENS:AVER2?
Password protection: Measure Protection
Sets the digital lter’s Filter Enable (see Section 5.7.3.3, DIGITAL FILTER (F3), on page 41).
The *RST command sets Filter Enable to the default setting which is ON.
8.7.3.85 SENS:AVER2:COUN <count>
Query form: SENS:AVER2:COUN?
Alternate parameter: DEF, MIN, MAX
Password protection: Measure Protection
Sets the digital lter’s Average Count setting (see Section 5.7.3.3, DIGITAL FILTER (F3), on page 41).
Parameter <count> is the average count. Allowed values range from 2 through 100.
This setting is only meaningful when Filter Enable is ON.
The *RST command sets Average Count to the default setting, which is 30.
8.7.3.86 SENS:AVER2:CLEA
Clears the digital lter buffer. This is equivalent to pressing the CLEAR FILTER function key in the Digital
Filter screen (see Section 5.7.3.3, DIGITAL FILTER (F3), on page 41).
8.7.3.87 SENS[<n>][:FRES]:DATA?
Example command: SENS4:DATA?
Example response: 44.221571
Reads a channel’s latest resistance measurement.
Sufx <n> is the channel number: 1 through 4 for the corresponding front channel input, 5 through 14 for
channels of the rst attached scanner, and 15 through 24 for channels of the second attached scanner. If the
sufx number is omitted, the latest measurement regardless of channel is returned.
If the Calculation setting for the channel is either RESISTANCE or TEMPERATURE (see Section 8.7.3.27,
CALC<n>:TYPE TEMP|RES|RAT, on page 80), a resistance value in ohms is returned. If calculation is
ratio, a resistance ratio value is returned. In either case, the returned value is obtained after processing by the
digital lter.
If no measurement is available, no response is returned and error code -230, “Data corrupt or stale” is gener-
ated in the error queue.
8.7.3.88 SENS[<n>]:RRAT:DATA?
Example command: SENS4:RRAT:DATA?
Example response: 1.30244715
Reads a channel’s latest resistance ratio measurement.
Sufx <n> is required, which species the channel number: 1 through 4 for the corresponding front channel
input, 5 through 14 for channels of the rst attached scanner, and 15 through 24 for channels of the second at-
tached scanner. If the sufx number is omitted, the latest measurement regardless of channel is returned.