Fluke 1595A Thermometer User Manual


 
1594A/1595A Super-Thermometer
Interface Commands
92
Parameter <value> is the number of measurements performed on a channel before advancing to the next chan-
nel. The range is 1 through 100 and the default is 1.
The *RST command sets Repeat Count to 1.
8.7.3.79 ROUT:SCAN:STAT <boolean>
Query form: ROUT:SCAN:STAT?
Alternate parameter: DEF
Sets the Scan Enable setting (see Section 5.7.3.1, SCAN SETTINGS (F1), on page 39).
The *RST command sets Scan Enable to the default setting which is ON.
8.7.3.80 SENS:APER[:DISP] <time>
Query form: SENS:APER[:DISP]?
Alternate parameter: DEF, MIN, MAX
Password protection: Measure Protection
Sets the measurement or Display Interval (see Section 5.7.3.2, TIMING SETTINGS (F2), on page 39).
Parameter <time> is in seconds. Allowed values for <time> are 1, 2, 10, 30, and 60. Values other than these
are rounded down to the nearest allowed value.
The Sample Period setting will be changed if necessary so that it is never greater than the Display Interval
setting.
The *RST command sets Display Interval to the default setting which is 2.
8.7.3.81 SENS:APER:PRES NORM|PREC|FAST|LONG
Query form: SENS:APER:PRES?
Alternate parameter: DEF
Sets the measurement timing settings to preset values (see Section 5.7.3.2, TIMING SETTINGS (F2), on
page 39). This is equivalent to pressing the PRESET NORMAL, PRESET PRECISION, PRESET FAST, or
PRESET LONG function key in the Timing Settings screen. It affects the Sample Period, Display Interval, and
Average Enable settings.
For the query form, if the timing settings match one of the presets, the name of the preset is returned. If the
settings do not match any of the presets, “NONE” is returned.
The *RST command sets the timing settings to the default preset which is NORM.
8.7.3.82 SENS:APER:SAMP <time>
Query form: SENS:APER:SAMP?
Alternate parameter: DEF, MIN, MAX
Password protection: Measure Protection
Sets the measurement Sample Period (see Section 5.7.3.2, TIMING SETTINGS (F2), on page 39).
Parameter <time> is in seconds. Allowed values for <time> are 1, 2, 5, and 10. Values other than these are
rounded down to the nearest allowed value.
The Display Interval setting will be changed if necessary so that it is never less than the Sample Period setting.
The *RST command sets Sample Period to the default setting which is 2 seconds.
8.7.3.83 SENS:AVER1[:STAT] <boolean>
Query form: SENS:AVER1?