Fluke 1595A Thermometer User Manual


 
53
Menus and Screens
Main Menu
5.7.5.3.1 STORE CONFIG (F1)
MAIN MENU | SYSTEM MENU | CONFIG |
The Store Conguration function allows the user to store settings to a conguration le in either internal non-
volatile memory or a USB ash memory device. The following settings are stored:
The probe denition assigned to each channel
Enable state of each channel
Channel settings, including the reference resistor selections
Resistor denition assigned to each reference resistor input
Scan settings
Timing settings
Digital lter settings
User Settings
Statistics elds settings
Statistics settings
Graph settings
Temperature unit
Recording settings
The Store Conguration function does not store probe denitions, resistor denitions, time and date, remote
interface port settings, password settings, calibration parameters, or recorded data.
MEMORY DEVICE - species to which memory device the conguration is stored. The options are USB
and INTERNAL. The default option is INTERNAL. Up to 50 conguration les may be stored internally. The
number of les stored on a USB device is limited only by the storage space available.
FILE NAME - species the name of the conguration le. It is edited or entered by rst pressing the ENTER
key, which opens the Alpha-Numeric Interface. The le name may have up to 24 characters. The user must
press the SAVE (F3) function key when nished. If the le does not exist, it is created. If it already exists, the
contents of the le are replaced. The le name extension “.txt” is automatically added to the le name when
the le is created. Once the memory device and le name are selected, the user presses the STORE (F1) func-
tion key to proceed.
USB conguration les are written to a folder named “\159x\cong” on the USB memory device and have the
extension “.txt”. When writing the le to a USB memory device, a formatted USB memory device must be
inserted into the USB socket on the front panel.
5.7.6 RECALL CONFIG (F2)
MAIN MENU | SYSTEM MENU | CONFIG |
The Recall Conguration screen displays the conguration les that have been saved to memory. The location
of the le(s) must be selected using the MEMORY DEVICE eld. The options are INTERNAL and USB.
Once the memory location is selected, pressing ENTER or the CONTINUE (F1) function key will open the
list of conguration les. Using the Up/Down arrow keys, the conguration is selected. Pressing ENTER or
the CONTINUE (F1) function key opens the le. The user is prompted to continue before the changes are
implemented.
The Recall Conguration function does not affect probe denitions, resistor denitions, time and date, remote
interface port settings, password settings, calibration parameters, or recorded data. If an assigned probe deni-
tion no longer exists in the Probe Library when a conguration is recalled, the channel is set to UNDEFINED.
Likewise, a reference resistor input will be set to VARIABLE if the resistor ID is not found in the Resistor
Library.