Installation, Operation, and Maintenance Manual # 509252 Rev R 99
Replacing the RF Gasket Assembly
1. If you detect evidence of arcing on the back of the reflector caused by the RF
gasket assemblies (behind the reflector), then you will need to replace the two
RF gasket assemblies with the main reflector removed.
Figure 38. Gasket Installation with Cross-guide (Short) Gaskets
(I310M/I311M)
2. Remove the housing (see “I310M/I311M Irradiator Housing
Removal/Replacement” on page 100).
IMPORTANT: The extrusion and the RF gasket assembly are one
assembly, and should not be separated from each other. To replace the RF
gasket requires removal of the old extrusion as well as the old RF gasket.
3. Access the casting by placing the lamp on its side and gently sliding the casting
out of the housing until the eight extrusion screws are accessible. (There are four
screws on each side).
RF SCREEN (Bottom of Irradiator)
Mesh
Cross-guide
(short) Gasket