Fusion F300SQ Light Therapy Device User Manual


 
84 F300S/F300SQ Ultraviolet Lamp System
7. Align the reflector squarely over the opening. With gentle pressure and a rocking
motion, slide the reflector down in place. It should not bind against the end
reflectors as it goes into place. As a final motion, grasp the sides of the reflector
and gently force them together a small amount, then settle the reflector down
over the mounting surfaces (Figure 23). Start the six flat head screws in the
reflector but do not tighten. Tighten the six screws in the sequence 1-6 as shown
in Figure 24 to a snug position; finally tighten to 3.4 N-m (30 inch-lbs.) using a
torque controlled screwdriver. Repeat tightening the screws as illustrated and at
the specified torque.
Figure 23. Reflector Removal or Replacement (I300M)
Figure 24. Reflector Screw Tightening Sequence