Fusion F300SQ Light Therapy Device User Manual


 
50 F300S/F300SQ Ultraviolet Lamp System
Power Source
NOTE: If a line voltage greater than 230V and a frequency of 50 Hz is used, see
the “F300S Relay Coil Resistor Assembly Installation” on page 53 at the end of
this chapter.
System Power Connections
Figure 14. Power Supply Front Panel
Listed current ratings indicate current demands during normal full-power operation.
Your customer-supplied wiring and circuit breaker or fuses should be sized to allow
for heavier current draw at startup. Current ratings are the same for 50 Hz or 60 Hz
systems. The recommended power cord to wire to P101 is:
CAUTION: Measure the main single-phase voltage
of your plant facilities to ensure that it conforms to
the factory-wired operating voltage specified on the
equipment label (rear panel of power supply).
Available line voltage should be within 10% of the
nominal operating voltage.
Fusion UV equipment must be connected in
accordance with local wiring codes. Use applicable
standards for wire sizes, fuse sizes, and disconnect
box locations. Each power supply requires a
separate, fused single-phase circuit with a line
disconnect.
F3q0004