Fusion F300SQ Light Therapy Device User Manual


 
Installation, Operation, and Maintenance Manual # 509252 Rev R 85
CAUTION: Do not use the flathead screws to pull the reflector into place.
Excessive pull on these screws will deform the reflector face at its bend
radius or loosen its rivets. This would result in a failure of the screen
assembly to seat properly and subsequent damage to the screen gaskets.
Figure 25. Correct Reflector Alignment (I300M)
8. When the reflector is properly installed, the face should be flush and even with
the outer face of the two end reflectors as shown in Figure 25.
9. Clean the reflector to remove fingerprints.
10. Clean and insert the lamp bulb according to “Cleaning Guidelines” on page 79.