Fusion F300SQ Light Therapy Device User Manual


 
108 F300S/F300SQ Ultraviolet Lamp System
Diode Test
An ohmmeter can be used to check for a shorted or open capacitor. (Isolate the
component before making continuity tests.)
Ohmmeters having less than a 6V battery can give erroneous indications of high
resistance in both directions when testing these high voltage diodes.
1. With an insulated screwdriver, short the capacitor terminals to each other to
assure discharge of capacitor voltage.
2. Disconnect the wiring from the terminals of the rectifier assembly.
3. Set the ohmmeter to highest resistance setting and measure resistance between
the AC terminals and both the - (minus) terminal and the GND terminal.
4. Measure the resistance between diode terminals with meter leads reversed.
Readings should be 50,000 ohms or less one way and infinite with leads
reversed.
WARNING: Disconnect all power sources
before servicing high voltage components.