Fusion F300SQ Light Therapy Device User Manual


 
94 F300S/F300SQ Ultraviolet Lamp System
CAUTION: Do not apply bending force to lamp bulb. Do not touch the
bulb with bare fingers or allow it to contact dirty or greasy surfaces.
Contaminants can be burned into the quartz envelope, causing
premature bulb failure.
Figure 32. Removing or Replacing the Bulb (I310M/I311M)
3. Clean any deposits from the surface of the bulb (see “Cleaning Guidelines” on
page 91).
4. To replace a lamp bulb in the reflector, hold it near the middle and insert the tip
of the bulb into the retaining hole of one end reflector. Apply pressure in the
direction of the length of the bulb, moving it toward the end that is inserted. This
will allow the free end to clear the end reflector. Gently slide the end of the bulb
down to the retaining hole and allow it to rest in position.
Reflector Removal/Replacement
Replace the reflector only if it becomes dull and its surface cannot be cleaned
successfully or if severe arcing is apparent.
1. Remove the screen assembly and the bulb.
Retaining hole