11 Tap the side of the res-
ervoir to remove any air bub-
bles.
12 To purge air bubbles that have risen
to the top of the reservoir, push up on the
plunger until you see insulin in the tub-
ing.
13 Without pulling, turn the
plunger counter-clockwise to re-
move it from the reservoir.
Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind your pump, and fill the infusion
set with insulin.
1
Remove the entire infusion set from your body.
2
If attached, remove the activity guard.
3
Turn the tubing connector half-turn counter
clockwise, then pull the reservoir and connector out
from the pump.
4
Safely dispose of the used reservoir and infusion set
item in a sharp's container.
5
You must now rewind your pump as described in the
next section.
tubing connector
half-turn counter-clockwise
56 Starting on insulin