Energizer 515 Blood Glucose Meter User Manual


 
AUTO OFF
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature, and so
insulin delivery has been stopped.
BATT OUT LIMIT
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and date are
correct. If not correct, go to the UTILITIES MENU and reset the date and time.
BOLUS STOPPED
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually
delivered. If necessary, reprogram a bolus with the amount not yet delivered.
BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/date.
E (ERROR)
After receiving this alarm, note the error number then call our 24 Hour HelpLine for assistance. An error alarm will
show the letter E followed by two numbers. E-alarms cause all insulin delivery to stop, the pump resets, and all
your settings are cleared.
EMPTY RESERVOIR
There is no insulin in the reservoir. Change the reservoir immediately.
Troubleshooting and alarms 133