Energizer 515 Blood Glucose Meter User Manual


 
Percent temp basal
Default value: 100% of basal programming
Power supply
The pump is powered by a standard 1.5 V AAA alkaline battery, size E92, type LR03 (Energizer brand recommended).
Prime function
Fixed prime range: 0.1 - 25.0 units (limited by maximum bolus)
Manual prime limit: Warning at 30 units, then at each 10 units thereafter.
Fill rate: 1 to 5 units/second.
Prime history
Maximum records shown: 20 manual and fixed
Program safety checks
Maximum infusion with single fault condition: 0.0 units
Pump size
The dimensions of the pump are approximately:
515 Pump:
2.0 x 3.0 (2.8 at the battery cap) x 0.75 inches
5.1 x 7.6 (7.1 at the battery cap) x 2.0 cm
715 Pump:
2.0 x 3.7 (3.5 at the battery cap) x 0.75 inches
5.1 x 9.4 (8.9 at the battery cap) x 2.0 cm
Pump specifications 145