Selecting temp basal type
Your pump will remember the temp basal type setting. Once the type is set, you do not have to set it again. To
select a temp basal type, do these steps:
1
Go to the SET TEMP BASAL AS screen.
Main > Basal > Temp Basal Type
2
The SET TEMP BASAL AS screen appears. Select Insulin Rate (U/H) or Percent of Basal, then press ACT.
3
The screen will return to the BASAL SETUP screen. The temp basal type is now set. Exit the menus.
If you select Percent of Basal as your temp basal type, changes to your basal rate are not allowed until after
temp basal is completed or cancelled.
Delivering a temp basal
A temp basal cannot exceed your programmed max basal rate.
1
Go to the BASAL MENU.
Main > Basal
2
Select Set/Edit Temp Basal, then press ACT.
3
The SET DURATION screen appears. The duration will flash. Duration is the amount of time it will take for the
pump to deliver the temporary basal. Enter the desired minutes or hours (30 minutes to 24-hours), then press
ACT.
4
If you have selected insulin rate as your temp basal type, the SET TEMP BASAL U/H screen appears. If you have
selected percent of basal as your temp basal type, the SET TEMP BASAL % screen appears. The temporary basal
rate will flash. Enter your temp basal rate, then press ACT.
5
The BASAL MENU screen appears with an open circle at the top of the screen. Your pump is in Special mode;
temp basal is now set and delivering. Exit the menus.