Energizer 515 Blood Glucose Meter User Manual


 
User settings
The user settings function allows you to save, restore, and clear all pump settings. You can also view a listing of
the dates and times of all recent user settings operations you have done. The Save Settings feature lets you keep
a set of pump settings that you can restore to your pump if it is cleared or you need to go back to these settings
for any reason.
When you clear your pump, the pump settings are restored to the factory defaults, and you must either use Restore
Settings, if you have saved a set of pump settings, or reprogram all your settings before you can use your pump
again. The pump does not clear the internal pump memory.
WARNING: Do not clear your pump settings while it is connected to your body.
CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a Medtronic
Diabetes representative. If you clear your pump settings, it will be necessary to reprogram all your personal
pump settings as directed by your healthcare professional. Additionally, you will have to rewind your pump.
Saving the settings
Do these steps to save your current pump settings:
1
Go to the UTILITIES MENU screen and select User Settings.
Main > Utilities > User Settings
2
Hold down and press ACT.
3
The USER SETTINGS screen is displayed with Save Settings highlighted. Press ACT.
4
If this is the first time you have saved pump settings, go to step 5. If you have previously saved pump settings,
a message displays, indicating the date of your last pump settings save. Read the instructions on the screen,
then press ACT to save your current settings. You can press ESC if you want to cancel the save.
5
The SETTINGS SAVED message displays to confirm that your current pump settings have been saved. Exit the
menus.
Restoring the settings
Do these steps to restore the most recent pump settings you have saved to your pump.
1
Go to the UTILITIES MENU screen and select User Settings.
Main > Utilities > User Settings