2
Hold down and press ACT.
3
The USER SETTINGS screen is displayed. Select Restore Settings, then press ACT.
4
A message displays, giving you the option of restoring the pump settings that were saved on the given date and
erasing the ones currently on your pump. Read the instructions on the screen then press ACT to restore the
settings. You can press ESC if you want to cancel the restore.
5
The SETTINGS RESTORED message displays to confirm that your current pump settings have been replaced with
the pump settings you had saved on the given date. Exit the menus and check your pump settings to verify the
restore.
Clearing the settings
Take the following steps only if you want to clear your pump to factory default settings.
WARNING: Do not clear your pump settings while it is connected to your body.
CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a Medtronic
Diabetes representative. If you clear your pump settings, it will be necessary to reprogram all your personal
pump settings as directed by your healthcare professional. Additionally, you will have to rewind your pump.
1
Go to the UTILITIES MENU screen and select User Settings.
Main > Utilities > User Settings
2
Hold down and press ACT.
3
The USER SETTINGS screen is displayed. Select Clear Settings, then press ACT.
4
A CONFIRM message displays, giving you the option of clearing your pump settings or not. Select YES and press
ACT to clear the settings. You can press ESC if you want to cancel.
5
You will see a RESET screen, and then the pump will go through various screens while it restarts. After the
pump clears all of your settings, the screen will go to the TIME/DATE SETUP screen.
6
Reset the time and date as described in Setting the time and date section of the Basic programming chapter.
7
After you set the time and date, you must rewind your pump. Refer to the Rewinding your pump section of the
Starting on insulin chapter for instructions. Remember, all your settings have been cleared and you must either
restore or reprogram all your settings.