Energizer 515 Blood Glucose Meter User Manual


 
FAILED BATT TEST
The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not used. If
the battery does not have enough voltage, this alarm will occur. The pump will not function and the battery must
be replaced. (Always make sure that you install a NEW battery into the pump.)
MAX DELIVERY
This alarm alerts you when you have taken more insulin than expected based on maximum bolus and maximum basal
rates.
MOTOR ERROR
Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.
NO DELIVERY
Insulin delivery has stopped. This alarm occurs if your pump detects a blockage.
NO RESERVOIR
The reservoir is not inserted correctly or no reservoir has been inserted.
OFF NO POWER
The battery is dead. Replace battery immediately. Follow the directions on the screen. Check to make sure that
the time is correct on the screen. Reset the time if necessary.
RESET
The Reset alarm triggers when pump settings are cleared because of one of these reasons:
Pump settings were cleared (Clear Settings function) and settings have not been reprogrammed.
A download attempt from the PC is incomplete. (The download function is applicable to the optional software
feature. Refer to the software user guide for more information.)
134 Troubleshooting and alarms