Energizer 515 Blood Glucose Meter User Manual


 
#5: Previous bolus activity results in a calculation of 1.5 units unabsorbed (active) insulin. User inputs 60 grams of
carbs and a BG of 200 mg/dL (11.1 mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
- 1.5 units (active insulin)
- 1.5 units (active insulin)
= 0.5 units+
(correction estimate)
= 2 + 0.5
= 2.5 units
bolus estimate
11.1 mmol/L - 6.7 mmol/L
2.2 mmol/L/u
= 0.5 units
200 mg/dL - 120 mg/dL
40 mg/dL/u
#6: Previous bolus activity results in a calculation of 3.5 units unabsorbed (active) insulin. User inputs 60 grams of
carbs and a BG of 200 mg/dL (11.1 mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
- 3.5 units (active insulin)
- 3.5 units (active insulin)
= -1.5 units*+
(correction estimate)
= 2 + 0
= 2 units
bolus estimate
11.1 mmol/L - 6.7 mmol/L
2.2 mmol/L/u
= -1.5 units*
200 mg/dL - 120 mg/dL
40 mg/dL/u
*This negative number indicates that active insulin is sufficient to cover the
correction that is needed. Thus, correction will be 0 units. Active insulin does
not reduce the food portion of the estimate.
Pump specifications 153