Tektronix 070-5992-05 Sleep Apnea Machine User Manual


 
Maintenance
4–8
760A / 760D / 760N
Table 4–1: Power Supply Fault Symptoms
Symptom Procedure
Line fuse open Rectifier/Switcher Check (Low Volts)
Power supply cycles OFF/ON Output Check (Low Volts), or
High Voltage Oscillator Check (High Volts)
Does not power up Control Circuit Check (Low Volts)
5 V not regulating Error Amplifier Check (Low Volts)
Improper CRT display High Volts Supply
NOTE. A 20W, 2-watt resistor should be used as a load for the Low Volts Supply.
Disconnect J4 and connect the 20W resistor between W1 (+5 V) and TP1
(secondary ground).
1. Preliminary Checks
a. A properly functioning and loaded Low Volts supply will output the
voltages listed in Table 4–2. Use the DMM to measure the voltages
between TP1 and the voltage test points. If the supply is not regulating
properly, continue with the procedure.
Table 4–2: Low Volts Supply Voltages
Test Point Voltage
W1 – (+5 V) +4.88 to +5.12 V
W4 – (+15 V) +14.0 to +16.0 V
W3 – (–15 V) –14.0 to –16.0 V
W2 – (+40 V) +39.0 to +41.0 V
NOTE. The Low Volts Power Supply troubleshooting is performed without
applying AC power.
b. Disconnect AC power from the instrument. Disconnect the instrument
from the Power Supply by removing the jumper from J4.
c. Use the digital multimeter to measure the voltage between TP2 and the
tab (drain) of Q9. Be sure the voltage is near 0 V before proceeding.
Low Volts Supply