Invacare TerminatorTM SS Mobility Aid User Manual


 
Part No. 1052712 Rev D 37 Terminator SS
FIGURE 13 - ADJUSTING WHEELBASE LENGTH
(ADJUSTING CENTER OF GRAVITY)
Indexing Marks (MUST be
set to Same Mark on
BOTH Suspension Tubes)
NEVER OPERATE THE
WHEELCHAIR WITH BOTH
ENDS OF THE INDEXING
TAPE SHOWING.
WHEELS SECTION 5
ADJUSTING WHEELBASE LENGTH (ADJUSTING CENTER
OF GRAVITY) (FIGURE 13)
WARNING
ALWAYS perform this procedure in the presence of an assistant. The position of
the footrest, camber tube, back angle, the tautness of the back upholstery as well
as the user's condition are directly related to the wheelchairs stability. Any
change to one (1) or any combination of the five (5) may cause the wheelchair to
decrease in stability. Use EXTREME caution when using a new seating position.
The addition of anti-tippers may be required .
1. Open the receiver tube clamps. Refer to OPENING/CLOSING CLAMPS in this section of the
manual.
WARNING
LENGTHENING THE WHEELBASE - NEVER position the camber system with
the end of the indexing tape (approximately 4-inches) of suspension tube
showing. The suspension tubes WILL NOT be securely tightened in the
wheelchair frame resulting in possible injury to the user and/or damage to the
wheelchair. Refer to FIGURE 13.
2. Position the camber system to the desired position on the wheelchair frame.
LENGTHENING THE WHEELBASE will increase the stability and maintain standard
maneuverability of the wheelchair. Make sure the end of the indexing tape is not showing.
SHORTENING THE WHEELBASE will decrease the stability, increase the
maneuverability and distribute additional weight onto the rear wheels.
NOTE: The performance of the wheelchair will be affected if the two (2) suspension tubes are not set to the
same indexing mark. Refer to FIGURE 13.
WARNING
QUICK RELEASE LEVERS - Make sure the quick release levers are
pointing towards the inside of the wheelchair and is in the CLOSED
position BEFORE using the wheelchair, otherwise personal injury or damage to
the wheelchair may result.
STANDARD - Make sure the hex screws and locknuts are securely
tightened BEFORE using the wheelchair, otherwise personal injury or damage to
the wheelchair may result.
3. Close the receiver tube clamps. Refer to OPENING/CLOSING CLAMPS in this procedure of
the manual.
4. Roll the wheelchair before using to make sure there is no excessive drag to either side.
NOTE: If drag to either side occurs, refer to DETERMINING TOE IN/TOE OUT in this section of the manual.
WHEELS