LogTag Recorders LogTag Analyzer Temperature Recorder Thermometer User Manual


 
Chapter 4 Preparing LogTags for use 39
When the Alert (Red) light on the LogTag flashes it means that a preset alert condition
has been exceeded. These configuration options allow you to determine how many
readings above or below selected limits are required for an alert on the LogTag to be
triggered. These can be set for consecutive readings beyond the limits and/or total
readings beyond the limit. So, if the red light is flashing, its meaning is based on how the
LogTag has been programmed.
For example, we may have configured a LogTag to flash the alarm if there were 5
continuous out-of-range readings or 7 total out-of-range readings.
This would yield the following response:
What happened Result
2 out-of-range, return to normal, 2 out-of-
range, return to normal, 2 out-of-range,
return to normal
No Flashing Red Alarm Light
4 out-of-range, return to normal, 4 out-of-
range
Flashing Red Alarm Light
4 out-of-range, return to normal, 2 out-of-
range
No Flashing Red Alarm Light