Lifescan UltraSmart Blood Glucose Meter User Manual


 
51
MESSAGE WHAT IT MEANS
Error message that
indicates the meter
has detected a
problem with the test
strip. Possible
causes are test strip
damage or an
incompletely filled
confirmation window.
The battery sign
appears on the
display. The power
of the batteries is
getting low. The
backlight will be
disabled when the
battery sign is
displayed. You can
complete about 100
more tests from the
time this symbol
first appears.
The power of the
batteries is too low
to run a test.
A
CTION
Repeat the test with
a new test strip.
Refer to pages 30–31
for information on
sample application.
Test results will still
be accurate, but
replace the batteries
as soon as possible.
Replace the
batteries at once.
WARNING
See Owner's Booklet
Error 5
Invalid test
Please repeat test
WARNING
Low Batteries !
You should replace
the batteries soon
See Owner’s Booklet
+
WARNING
Batteries !
You MUST replace
the batteries now
See Owner’s Booklet
+
DCO_06158108A_OUS_OB_GB 5/6/03 1:25 PM Page 51