Emerson Process Management 3051N Sleep Apnea Machine User Manual


 
Reference Manual
00809-0100-4808, Rev CA
June 2008
A-17
Rosemount 3051N
ROM checksum
error ON
Checksum of transmitter software has detected a fault. Return
to Rosemount Nuclear Instruments, Inc. for repair.
Save data from
device to
configuration
memory
Prompts user to press SAVE softkey to initiate a
device-to-memory transfer.
Saving data to
configuration
memory.
Data is being transferred from a device to configuration
memory.
Sending data to
device.
Data is being transferred from configuration memory to a
device.
Sensor board not
initialized ON
The sensor module electronics board is not initialized. Return
to Rosemount Nuclear Instruments, Inc. for repair.
There are write only
variables which
have not been
edited. Please edit
them.
There are write-only variables which have not been set by the
user. These variables should be set or invalid values may be
sent to the device.
There is unsent data.
Send it before
shutting off?
Press YES to send unsent data and turn the HC off. Press NO
to turn the HC off and lose the unsent data.
Too few data bytes
received
Command returns fewer data bytes than expected as
determined by the device description.
Transmitter Fault Device returns a command response indicating a fault with the
connected device.
Units for <variable
label> has changed.
Unit must be sent
before editing, or
invalid data will be
sent.
The engineering units for this variable have been edited. Send
engineering units to the device before editing this variable.
Unsent data to
online device. SEND
or LOSE data
There is unsent data for a previously connected device which
must be sent or thrown away before connecting to another
device.
Upgrade 275
software to access
XMTR function.
Continue with old
description?
The communicator does not contain the most recent
Rosemount 3051N Device Descriptors (DDs). Select YES to
communicate using the existing DDs. Select NO to abort
communication.
Use up/down arrows
to change contrast.
Press DONE when
done.
Gives direction to change the contrast of the HC display.
Message Description