Emerson Process Management 3051N Sleep Apnea Machine User Manual


 
Reference Manual
00809-0100-4808, Rev CA
June 2008
A-15
Rosemount 3051N
Add item for ALL
device types or only
for this ONE device
type.
Asks the user whether the hot key item being added should be
added for all device types or only for the type of device that is
connected.
Command Not
Implemented
The connected device does not support this function.
Communication
Error
The communicator and the device are not communicating
correctly. Check all connections between the communicator
and the device and resend the information.
Configuration
memory not
compatible with
connected device
The configuration stored in memory is incompatible with the
device to which a transfer has been requested.
CPU board not
initialized ON
The electronics board is not initialized. Return to Rosemount
Nuclear Instruments, Inc. for repair.
CPU EEPROM write
failure ON
Message sent to electronics board from HART signal failed.
Return to Rosemount Nuclear Instruments, Inc. for repair.
Device Busy The connected device is busy performing another task.
Device
Disconnected
The device failed to respond to a command. Check all
connections between the communicator and the device and
resend the command.
Device write
protected
Device is in write-protect mode. Data can not be written.
Device write
protected. Do you
still want to
shut off?
Device is in write-protect mode. Press YES to turn the HART
communicator off and lose the unsent data.
Display value of
variable on hotkey
menu?
Asks whether the value of the variable should be displayed
adjacent to its label on the hotkey menu if the item being added
to the hotkey menu is a variable.
Download data from
configuration
memory to device
Press the SEND softkey to transfer information from the
communicator memory to the device.
Exceed field width Indicates that the field width for the current arithmetic variable
exceeds the device-specified description edit format.
Exceed precision Indicates that the precision for the current arithmetic variable
exceeds the device-specified description edit format.
Ignore next 50
occurrences of
status?
Select YES to ignore the next 50 occurrences of device status,
or select no to display every occurrence.
Illegal character An invalid character for the variable type was entered.
Illegal date The day portion of the date is invalid.
Illegal month The month portion of the date is invalid.
Illegal year The year portion of the date is invalid.
Incompatible CPU
board and module
ON
Return to Rosemount Nuclear Instruments, Inc. for repair.
Incomplete
exponent
The exponent of a scientific notation floating point variable is
incomplete.
Incomplete field The value entered is not complete for the variable type.
Message Description