Agilent Technologies 1142A Thermometer User Manual


 
51
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
7 Connect the adapter/probe combination to the test board in the position shown
in the figure below.
Figure 2-23
Signal to + input
8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top.
Use the figure below for adjust locations.
Figure 2-24
Adjustment Locations
9 Change the function generator to 10 V
p-p
(10x adapter adjustment only).
10 On the oscilloscope, press CHAN and set the sensitivity to 1 mV/div.