Agilent Technologies 1142A Thermometer User Manual


 
39
Calibration Tests and Adjustment
Calibration Test Record
Calibration Test Record
1141A/1142A Differential Probe Tested by_________________________
Serial No. ______________________________ Work Order No.____________________
Recommended Test Interval - 1 Year/2000 hours Date____________________
Recommended next testing_________________ Temperature_____________
Test Limits Results
dc Gain Accuracy Probe
Only
+0.98 mV to +1.02 mV _____________
10x +0.096 mV to +0.104 mV _____________
100x +0.0096 mV to 0.0104 mV _____________
Bandwidth >0.707 at 200 MHz _____________
CMRR 1 MHz 3000:1 _____________
100 MHz 10:1