Agilent Technologies 1142A Thermometer User Manual


 
33
Calibration Tests and Adjustment
dc Gain Accuracy
6 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to - input).
Figure 2-7
Signal to - input
7 Record the V
out2
measurement from the bottom DVM in figure 2-2.
8 Calculate the 100x attenuator gain as
Record the result of this calculation in the “Calibration Test Record” on page 39
NOTE Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment
of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter
Adjustment procedure in the Adjustments section later in this chapter then retest the
attenuator adapter. If if continues to fail, repair is necessary.
V
out
V
in
--------------
V
out
1
V
out
2
()
2 V
in
1
×
----------------------------------=