6. Assembly
Transmitted light and incident light analyzer
Remove the plug cap on the left side of the
stand.
Insert the analyzer into the receptacle until it
latches in place (Fig. 30).
Motorized analyzer
Insert the analyzer cube into the correspond-
ing position on the filter turret as described in
section "6.7 Equipping the incident light turret
disk" p. 30. For details on the proper
position for the analyzer cube, please refer to
the identification sheet provided.
Fig. 30 Inserting the analyzer
1 The plug cap is replaced with the analyzer.
6.9 DIC prisms
Insert the objective prism slide into the tube
slot (Fig. 31.1). The code letter must match the
code letter on the objective.
In the Leica DM5000 B microscope, the DIC
prisms are already installed in the DIC disk
above the objective turret (Fig. 68).
Fig. 31 Inserting the objective prism slide
1 Objective prism slide
1