Leica DM4000M Microscope & Magnifier User Manual


 
30
6. Assembly
Fig. 25 Removing the front panel
1 Filter receptacle
2 Locking pin
3 Front panel
Fig. 26 Inserting the filter or reflector cubes
1 Mounting
1
2
Fig. 23 Filter cube,
front side
6.7 Equipping the incident light turret disk
The positions in the turret disk are numbered.
Depending on how they are equipped, the indi-
vidual filter and/or reflector cubes are set in
pre-assigned positions at the factory. For
details, check the identification sheet included
with your order.
Insert the filter and reflector cubes in the fol-
lowing manner:
Never fit the incident light turret disk while
the microscope is in operation.
Remove the face plate from the upper part of
the microscope (Fig. 25). Press the locking pin
(25.2) to turn the turret disk. When the locking
pin is released, the turret disk locks into place
again.
With the holder facing you squarely, insert the
filter cube or reflector cube into the holder as
described in the identification sheet provided.
To do so, place the filter or reflector cube on
the right side and press it toward the left into
the mounting (Fig. 26).
Press the locking pin (25.2) and turn the filter
turret to the next click stop.
Make sure that the turret engages (the lock-
ing pin springs forward) and insert the next fil-
ter and/or reflector cube as described above.
When all filters and reflector cubes have been
inserted, close the front cover plate again.
1
3
1
Fig. 24 Filter cube,
back side