Leica DM4000M Microscope & Magnifier User Manual


 
31
6. Assembly
6.8 Polarizer and analyzer
Transmitted light polarizer: ICT/P
Using the left clamping screw, fasten the ICT/P
transmitted light polarizer to the underside of
the condenser holder (Fig. 27).
Make sure that the red index point on the
front of the polarizer is at 0.
If necessary, insert the compensators (λ-, λ/4
plates or elliptical compensators) into the po-
larizer’s receptacle (Fig. 28) or into the com-
pensation slot.
Incident light polarizers:
R/P polarizer, rotating polarizer,
L/ICR polarizer, R/ICR polarizer
Remove the plug cap on the right side of the
incident light axis (Fig. 29).
Insert the polarizer into the receptacle until it
locks into place.
Caution:
Push the polarizer into the front receptacle only.
Motorized polarizer
A motorized polarizer is already installed and
ready for operation in the DIC condenser.
Fig. 27 Assembly of the ICT/P transmitted light polarizer
1 Clamping screw
Fig. 28 Inserting the compensators
Fig. 29 Inserting the polarizer
1 The plug cap is replaced with the polarizer.