Invacare 1073955 Mobility Aid User Manual


 
SECTION 6—ARMS
Orbit
50 Part No. 1073955
Replacing/Repositioning the Locking Mechanism in the
Cantilever Arm
NOTE:Forthisprocedure,refertoFIGURE 6.10onpage50.
CAUTION
The locking mechanism is spring loaded. Place your free hand over the locking
mechanism to prevent parts from springing out of the armrest.
1. Movethecantileverarmupandoutoftheway.
2. Unthreadtheactuatorfromthelockingmechanismandremove.
3. Slowlyletthelockingmechanismandspringslideoutofthecantileverarm.
NOTE:Inspectthespringforwearanddamage.Replaceifnecessary.
4. Slidethenewlockingmechanismandspringintothecantileverarm.
5. Positiontheangledportionofthelockingmechanisminoneoftwoways:
•AngledPortionFacingUp‐Armwillflipup
•AngledPortionFacingDown‐Armwillflipdown
6. UseLoctite242andsecurelytightentheactuatorintothelockingmechanism.
7. Tolockthecantileverarm,pushdown(orpullup)untilthereisanaudibleclick.
8. Pullup(orpushdown)onthecantileverarmtomakesureitislockedinplace.
FIGURE 6.10 Replacing/Repositioning the Locking Mechanism in the Cantilever Arm
Angled Portion of
Locking Mechanism
Cantilever Arm
Actuator
(Apply Loctite 242)
Spring
Locking
Mechanism