Invacare 1073955 Mobility Aid User Manual


 
SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
Part No. 1073955 25 Orbit
Troubleshooting
Maintenance
Maintenance Safety Precautions
ƽ WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely.
CAUTION
DO NOT overtighten hardware attaching to the frame. This could cause damage to
the frame tubing.
Suggested Maintenance Procedures
1. Beforeusingthewheelchair,makesureallnutsandboltsaretight.Checkallpartsfor
damageorwearandreplace.Checkallpartsforproperadjustment.
2. Keepquickreleaseaxlesfreeofdirtandlinttoensurepositivelockingandproper
operation.RefertoAdjustingtheQuickReleaseAxleonpage 62.
3. Oilquickreleaseaxlesatleastonceamonth(3in1™oilorequivalent).
4. NONRECLINERSEATFRAMES‐Periodicallycheckthebackfolddown
mechanismstoensurethattheylockthebacksecurelyinplace.Disassembleandclean
ifnecessary.RefertoReplacingtheLockingMechanismintheBackCaneonpage 52.
Chair
Veers
Right/
Left
Chair 3
Wheels
Sluggish Turn or
Performance
Casters
Flutter
Squeaks
and
Rattles
Looseness
in
Chair
Solutions
XX X X
Check tires for
correct and equal
pressure.
XXXX
Check for loose nuts
and bolts.
XX X
Check caster headtube
angle.
XX
Check that rear
wheels are equally
spaced away from seat
frame.