Emerson OCX 8800 Oxygen Equipment User Manual


 
Instruction Manual
IM-106-880, Rev 1.0
January 2007
6-25
OCX 8800
Figure 6-20. Installation of O
2
Cell
and Heater Strut Assembly
Install Sample Block Heater Rods
1. Before installing sample block heater rods (13, Figure 6-20), evenly coat
the heater rods with Watlube heater release agent.
5
6
12
11
14
15
10
9
1
2
4
3
7
8
13
3
7
390066
1. Screw
2. Locking Clip
3. Cover
4. O-ring
5. Marking Plate
6. Terminal Insulator
7. Reference Air Tube
8. Sensor Housing
9. Heater Strut Assembly
10. Gasket
11. Screw
12. Heater Clamp
13. Heater Rod
14. Insulator
15. Thermal Switch
15