Emerson OCX 8800 Oxygen Equipment User Manual


 
Instruction Manual
IM-106-880, Rev 1.0
January 2007
B-3
OCX 8800
IN IN
Fluke Model 87
Multimeter
or equivalent
Optional
mA
AC or DC
Power Input
AC
ACC
GND
37390024
250
950
>
<
OCX
8800
Figure B-3. SPA Jumper
and Dip Switch Settings
Figure B-4. SPA Setup for
Calibration
2. Verify that the Password Jumper is set to the OFF position. If the jumper
is in the ON position, reposition the jumper.
3. Check the position of the Failsafe Dip Switches. Position the dip
switches as shown in Figure B-3.
4. Check the position of the Source Current Dip Switches. Position the dip
switches as shown in Figure B-3.
5. Reinstall the SPA access cover.
Configuration/Calibration
Prior to operation, the SPA operating parameters must be configured via a
menu-driven setup procedure. At the end of the configuration procedure, the
SPA analog output signal is calibrated to insure valid communications.
1. See Figure B-4. Connect a calibrated ammeter (Fluke Model 87 or
equivalent, accurate to ±0.025%) to the SPA analog output terminals.
Observe polarity.
2. Connect a 90 to 260 VAC or 22 to 300 VDC power source to the SPA
power terminals. When connecting an AC power source, use the AC
and ACC (AC Common) terminals. For a DC source, use the AC and
Ground terminals.
3. If desired, you can connect the 4 to 20 mA O2 signal wires from the
OCX 8800 analog output terminal block to the SPA Input terminals. (The
OCX must be operational to transmit the O2 signal. Observe polarity.)
Access Cover
37020022
Password
ON
Password
OFF
Password
Jumper Pins
Failsafe Dip Switches
(shown in correct position)
(shown in correct position)
Source Current Dip Switches
SPA - Bottom View
NOTICE
REMOVETHIS PANEL
FOR SERVICE ACCESS.
STATIC SENSITIVE.
USE CAUTIONWHEN
CHANGING JUMPERS