Emerson OCX 8800 Oxygen Equipment User Manual


 
Instruction Manual
IM-106-880, Rev 1.0
January 2007
1-11
OCX 8800
Table 1-1. Product Matrix - Hazardous Area OCX 8800
OCX88C O
2
/Combustibles Transmitter - Flameproof
Code Probe Length and Material
00 No Probe or Exhaust Tube
11 18 in. (457 mm) 316 SST up to 1300°F (704°C)
21 18 in. (457 mm) Inconel 600 up to 2600°F (1427°C)
31 18 in. (457 mm) Ceramic up to 2600°F (1427°C)
12 3 ft (0.91 m) 316 SST up to 1300°F (704°C)
22 3 ft (0.91 m) Inconel 600 up to 1832°F (1000°C)
32 3 ft (0.91 m) Ceramic up to 2600°F (1427°C)
13 6 ft (1.83 m) 316 SST up to 1300°F (704°C)
23 6 ft (1.83 m) Inconel 600 up to 1832°F (1000°C)
14 9 ft (2.7 m) 316 SST up to 1300°F (704°C)
24 9 ft (2.7 m) Inconel 600 up to 1832°F (1000°C)
Code Probe Mounting Assembly
10 (ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes
20 (DIN) 185 mm dia. flange, 145 mm BC with 4 x 18 mm dia. holes
Code Mounting Hardware - Stack Side
0 No Adapter Plate (“0” must be chosen under “Mounting Adapter - Probe Side” below)
1 New Installation - Square weld plate with studs
2 Model 218/240 Mounting Plate (with Model 218/240 Shield Removed)
3 Existing Model 218/240 Support Shield
4 Special Mounting
(1)
5 Model 132 Adapter Plate
Code Mounting Hardware - Probe Side
0 No Adapter Plate
1 Probe Only (ANSI)
2 Probe Only (DIN)
Code Electronics Housing - Communications
H1 HART Communications
(2)
H2 HART Communications with LOI
H3 HART Communications with Calibration Solenoids
(2)
H4 HART Communications with LOI and Calibration Solenoids
Code Electronics Mounting
02 Remote Electronics and no cable
Code Accessories
00 None
01 Flow meters & Ref. Air Set
02 In-Situ Filter (Stainless Steel only)
03 In-Situ Filter (SST), Flow meters & Ref. Air Set
11 Flow meters, and Ref. Air Set with Blowback
12 In-Situ Filter (SST) with Blowback
13 In-Situ Filter (SST), Flow meters & Ref. Air Set with Blowback
OCX88C 11 10 1 1 H2 02 02 Example
NOTES:
(1)
Provide details of the existing mounting plate as follows:
Plate with studs Bolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate.
Plate without studs Bolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
(2)
If the LOI is not implemented, remote access and functionality available via HART Communications (Model 275/375 Handheld Communicator)
with Oxygen/Combustibles Device Description (DD) required.